DEVICE: Prometra II Pump (00810335020228)

Device Identifier (DI) Information

Prometra II Pump
13827
In Commercial Distribution

Flowonix Medical, Inc.
00810335020228
GS1

1
013518517 *Terms of Use
The Prometra II Programmable Pump is a sterile, battery-operated, teardrop-shaped implantable, programmable infusion pump, with a rigid titanium housing and flow controller system, which dispenses infusate into the intrathecal space through an implanted intrathecal catheter. To help increase safety, the Prometra II Pump incorporates a safety valve (flow-activated valve or FAV) that will shut off drug flow to the patient in the event a high flow rate occurs, such as during an MRI. All functions of the system (e.g., dosing) are controlled externally using a hand-held, battery-operated programmer. The Prometra II Pump contains a metal bellows drug reservoir with a capacity of 20 milliliters (mL). The reservoir propellant is stored within the rigid housing surrounding the bellows and provides the driving pressure for the pump. The driving pressure on the reservoir forces the infusate through an outlet filter (0.22 pm), and into an electronically controlled flow metering valve-accumulator subsystem. The Infumorph passes from the flow metering subsystem, into the catheter access port then into the catheter for delivery to the intrathecal space.
CLOSE

Device Characteristics

MR Conditional
No
Yes
Yes
Yes
No
No
No
No

GMDN

[?]

GMDN© Term Code, Names and Definitions (*Terms of Use): GMDN® is a registered trademark of The GMDN Agency. All rights reserved. Used under licence from The GMDN Agency Ltd.

GMDN Term Code GMDN Term Name GMDN Term Definition GMDN Term Status [?] Implantable?
46024 Implantable intrathecal infusion pump, programmable
A battery-powered programmable device designed to be implanted in a patient for the storing and subarachnoid (intrathecal) administration of narcotics/drugs (e.g., morphine sulfate, baclofen) to manage intractable pain and muscle spasms of malignant or non-malignant origin. This implantable infusion pump (IIP) delivers drug doses from its implanted refillable reservoir which is controlled by drug concentration and/or by radio-frequency (RF) signals from an external programming device. The drug reservoir, usually implanted under the skin of the lower abdomen, is typically connected to a catheter placed into the spinal fluid space.
Active true
CLOSE

FDA Product Code

[?]
Product Code Product Code Name
LKK Pump, Infusion, Implanted, Programmable
CLOSE

FDA Premarket Submission

FDA Premarket Submission Number [?] Supplement Number [?]
P080012 022
No CLOSE

Sterilization

Yes
No
Sterilization Method [?]
No Sterilization Methods Found
CLOSE

Storage and Handling

[?]
Storage and Handling
Storage Environment Temperature: between 2 and 52 Degrees Celsius
CLOSE

Clinically Relevant Size

[?]
Size Type Text
No Device Sizes
CLOSE

Device Record Status

acd6019a-7200-487f-87ba-90b86c83856c
April 03, 2024
5
April 29, 2015
CLOSE

Alternative and Additional Identifiers Additional Identifiers

Package DI

[?]
Package DI Number Quantity per Package Contains DI Package Package Discontinue Date Package Status Package Type
No Package DIs found
CLOSE

Secondary DI

[?]
Issuing Agency [?] Secondary DI Number
No Secondary DIs found
CLOSE

Unit of Use DI

[?]
Unit of Use DI Number: No Unit of Use DI Numbers Found CLOSE

Direct Marking (DM)

[?]
No
No
None
CLOSE

Production Identifier(s) in UDI

[?]
Yes
Yes
Yes
No
No CLOSE

Customer Contact

[?]
No Customer Contact currently defined
CLOSE