The Carl Zeiss Meditec MEL 90 Excimer Laser System is designed for refractive surgery based on the ablation of corneal tissue achieved with a short pulse excimer laser having a wavelength of 193 nanometers. The laser head emits 4 to 7 nanosecond pulses (FWHM nominal pulse duration) with a repetition rate of 500 Hz. The MEL 90 Excimer Laser is a spot-scanning laser that utilizes a Gaussian beam with a 0.7 mm spot diameter.
Indications for Use:
The MEL 90 is indicated for use in primary Laser-Assisted in situ Keratomileusis (LASIK) treatments for the reduction or elimination of:
• Myopia less in magnitude than -10.00 D sphere (in minus-cylinder notation), with and without cylinder up to -4.00 D, when MRSE is less in magnitude than -10.00 D;
• Hyperopia up to +4.00 D sphere (in plus-cylinder notation), with and without cylinder up to +3.00 D, when MRSE is up to +5.00 D; and
• Mixed astigmatism with cylinder from >1.00 D up to 4.00 D;
in patients who are 18 years of age or older with documentation of stable manifest refraction over the past year as demonstrated by a change in MRSE within +/- 0.50 D.
Device Characteristics
Labeling does not contain MRI Safety Information | |
No | |
No | |
No | |
Yes | |
No | |
No | |
No | |
No |
GMDN
[?]GMDN© Term Code, Names and Definitions (*Terms of Use): GMDN® is a registered trademark of The GMDN Agency. All rights reserved. Used under licence from The GMDN Agency Ltd.
GMDN Term Code | GMDN Term Name | GMDN Term Definition | GMDN Term Status [?] | Implantable? |
---|---|---|---|---|
17702 | Ophthalmic excimer laser system |
A mains electricity (AC-powered) device assembly in which input energy is used to excite an excimer gas to emit a high-power laser beam intended for corneal ablation (i.e., photorefractive keratectomy) and other ophthalmologic procedures (e.g., surgical creation of a communication between the lacrimal sac and the nasal cavity). It includes a light source and controls/foot-switch, and is typically coupled to a biomicroscope slit lamp or to an indirect ophthalmoscope by a flexible fibreoptic cable. It does not include femtosecond pulsing or fundus imaging technology.
|
Active | false |
FDA Product Code
[?]Product Code | Product Code Name |
---|---|
LZS | Excimer Laser System |
FDA Premarket Submission
FDA Premarket Submission Number [?] | Supplement Number [?] |
---|---|
P060004 | 006 |
Sterilization
Storage and Handling
[?]Storage and Handling |
---|
Handling Environment Temperature: between 18 and 24 Degrees Celsius |
Handling Environment Humidity: between 0 and 50 Percent (%) Relative Humidity |
Handling Environment Atmospheric Pressure: between 70 and 106 KiloPascal |
Storage Environment Temperature: between -15 and 40 Degrees Celsius |
Storage Environment Humidity: between 0 and 95 Percent (%) Relative Humidity |
Storage Environment Atmospheric Pressure: between 50 and 106 KiloPascal |
Clinically Relevant Size
[?]Size Type Text |
---|
Weight: 295 Kilogram |
Device Size Text, specify: x W x H: 1.63 m x 0.73 m x 1.48 m to 1.70 m |
Device Record Status
162179b1-b91d-42b7-be4d-00f613370095
December 23, 2024
1
December 13, 2024
December 23, 2024
1
December 13, 2024
Alternative and Additional Identifiers Additional Identifiers
Package DI
[?]Package DI Number | Quantity per Package | Contains DI Package | Package Discontinue Date | Package Status | Package Type |
---|---|---|---|---|---|
No Package DIs found |
Secondary DI
[?]Issuing Agency [?] | Secondary DI Number |
---|---|
No Secondary DIs found |
Unit of Use DI
[?]
Unit of Use DI Number:
No Unit of Use DI Numbers Found
CLOSE
Direct Marking (DM)
[?]Production Identifier(s) in UDI
[?]Customer Contact
[?]
No Customer Contact currently defined