DEVICE: ABL90 FLEX PLUS Analyzer (05700693930923)

Device Identifier (DI) Information

ABL90 FLEX PLUS Analyzer
393-092
In Commercial Distribution
393-092
Radiometer Medical ApS
05700693930923
GS1

1
308667539 *Terms of Use
ABL90 FLEX PLUS, Analyzer without casettes, cables, manuals etc.
CLOSE

Device Characteristics

Labeling does not contain MRI Safety Information
No
No
No
Yes
No
No
No
No

GMDN

[?]

GMDN© Term Code, Names and Definitions (*Terms of Use): GMDN® is a registered trademark of The GMDN Agency. All rights reserved. Used under licence from The GMDN Agency Ltd.

GMDN Term Code GMDN Term Name GMDN Term Definition GMDN Term Status [?] Implantable?
56667 Metabolic profile clinical chemistry analyser IVD
A portable, electrically-powered automated or semi-automated laboratory instrument intended to be used for the qualitative and/or quantitative in vitro measurement of multiple clinical chemistry analytes to generate a metabolic profile. Analytes may include pH, blood gas parameters (e.g., pO2, pCO2), electrolytes [e.g., Sodium (Na2+), potassium (K+), chloride (Cl-)], glucose, urea, creatinine and sometimes in addition haemoximetry parameters [e.g., carboxy-haemoglobin (COHb), methaemoglobin (MetHb), total haemoglobin (totHb), hematrocrit (Hct)].
Active false
CLOSE

FDA Product Code

[?]
Product Code Product Code Name
MQM BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM
KQI ASSAY, FETAL HEMOGLOBIN
KHP ACID, LACTIC, ENZYMATIC METHOD
JFP ELECTRODE, ION SPECIFIC, CALCIUM
CEM ELECTRODE, ION SPECIFIC, POTASSIUM
JGS ELECTRODE, ION SPECIFIC, SODIUM
CHL ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH
JIX CALIBRATOR, MULTI-ANALYTE MIXTURE
CGA GLUCOSE OXIDASE, GLUCOSE
GKR SYSTEM, HEMOGLOBIN, AUTOMATED
JJY Multi-analyte controls, all kinds (assayed)
CGZ ELECTRODE, ION-SPECIFIC, CHLORIDE
GHS ASSAY, CARBOXYHEMOGLOBIN
CLOSE

FDA Premarket Submission

FDA Premarket Submission Number [?] Supplement Number [?]
Premarket Submission Number Not Available/Not Released
No CLOSE

Sterilization

No
No
Sterilization Method [?]
No Sterilization Methods Found
CLOSE

Storage and Handling

[?]
Storage and Handling
Storage Environment Temperature: between -20 and 60 Degrees Celsius
Handling Environment Temperature: between 15 and 32 Degrees Celsius
Handling Environment Atmospheric Pressure: between 70 and 107 KiloPascal
Handling Environment Humidity: between 10 and 80 Percent (%) Relative Humidity
Special Storage Condition, Specify: Labeling does not contain MRI Safety information
CLOSE

Clinically Relevant Size

[?]
Size Type Text
No Device Sizes
CLOSE

Device Record Status

788523a7-263f-413c-8900-4ad2e6a84dd8
November 14, 2024
5
February 15, 2017
CLOSE

Alternative and Additional Identifiers Additional Identifiers

Package DI

[?]
Package DI Number Quantity per Package Contains DI Package Package Discontinue Date Package Status Package Type
No Package DIs found
CLOSE

Secondary DI

[?]
Issuing Agency [?] Secondary DI Number
No Secondary DIs found
CLOSE

Unit of Use DI

[?]
Unit of Use DI Number: No Unit of Use DI Numbers Found CLOSE

Direct Marking (DM)

[?]
No
No
None
CLOSE

Production Identifier(s) in UDI

[?]
Yes
No
No
Yes
No CLOSE

Customer Contact

[?]
+45999999999
radiometer@radiometer.dk
CLOSE