DEVICE: NA (08800019200253)
Device Identifier (DI) Information
NA
Anyview-240DR
In Commercial Distribution
ECO-ER-CAT-001DR
ECOTRON
Anyview-240DR
In Commercial Distribution
ECO-ER-CAT-001DR
ECOTRON
1. Line Voltage: 220~230V, 50/60Hz
2. Output Power: 2.4kW
3. Fluoroscopy Mode
- F-mode: 0.5~10mA
- P-mode: 0.5~20mA
- Boost mode: 30mA
- DS mode: 10~30mA
4. Radiography Mode
- kV range: 40~110kV, mA range: 16~30mA, mAs range: 0.8~100mAs
5. X-ray Tube
- XRR-2551(Rotating Type), Focal Spot: 0.3/0.6, Target Angle: 10degree, Anode Heat Storage: 210kHU
6. Detector
- VIVIX-D 1212G (VIEWORKS), 12 x 12 inch
Device Characteristics
Labeling does not contain MRI Safety Information | |
No | |
No | |
No | |
No | |
No | |
No | |
No | |
No |
GMDN
[?]GMDN© Term Code, Names and Definitions (*Terms of Use): GMDN® is a registered trademark of The GMDN Agency. All rights reserved. Used under licence from The GMDN Agency Ltd.
GMDN Term Code | GMDN Term Name | GMDN Term Definition | GMDN Term Status [?] | Implantable? |
---|---|---|---|---|
37646 | Mobile general-purpose fluoroscopic x-ray system, digital |
A mobile (within an imaging facility) general-purpose diagnostic fluoroscopic x-ray system that uses a C-arm and digital techniques for image capture, display and manipulation and is designed to be used in a variety of general-purpose applications requiring real-time fluoroscopic imaging capabilities. It includes spot-film capabilities in addition to the fluoroscopic features and is intended to optimize the capability of users to visually and quantitatively evaluate the anatomy and physiological function of various targeted body areas in real-time. It is frequently used in conjunction with an ingested or injected x-ray contrast medium. Images can be both real-time and delayed formats.
|
Active | false |
FDA Product Code
[?]Product Code | Product Code Name |
---|---|
OXO | Image-Intensified Fluoroscopic X-Ray System, Mobile |
JAA | System, X-Ray, Fluoroscopic, Image-Intensified |
OWB | Interventional Fluoroscopic X-Ray System |
FDA Premarket Submission
FDA Premarket Submission Number [?] | Supplement Number [?] |
---|---|
K180473 | 000 |
Sterilization
Storage and Handling
[?]Storage and Handling |
---|
Handling Environment Temperature: between 10 and 40 Degrees Celsius |
Handling Environment Humidity: between 30 and 75 Percent (%) Relative Humidity |
Storage Environment Temperature: between -10 and 70 Degrees Celsius |
Storage Environment Humidity: between 5 and 95 Percent (%) Relative Humidity |
Clinically Relevant Size
[?]Size Type Text |
---|
Weight: 250 Kilogram |
Width: 800 Millimeter |
Length: 1852 Millimeter |
Height: 1650 Millimeter |
Device Record Status
2b994469-d7aa-4454-886a-6c7928b92007
July 06, 2018
2
March 29, 2018
July 06, 2018
2
March 29, 2018
Alternative and Additional Identifiers Additional Identifiers
Package DI
[?]Package DI Number | Quantity per Package | Contains DI Package | Package Discontinue Date | Package Status | Package Type |
---|---|---|---|---|---|
No Package DIs found |
Secondary DI
[?]Issuing Agency [?] | Secondary DI Number |
---|---|
No Secondary DIs found |
Unit of Use DI
[?]
Unit of Use DI Number:
No Unit of Use DI Numbers Found
CLOSE
Direct Marking (DM)
[?]Production Identifier(s) in UDI
[?]Customer Contact
[?]
9999999999
jun1117@ecotron.co.kr
jun1117@ecotron.co.kr