DEVICE: LungFit PH NO2 Filter (B882900080)

Device Identifier (DI) Information

LungFit PH NO2 Filter
90008
In Commercial Distribution

Beyond Air, Inc.
B882900080
HIBCC

1
080243711 *Terms of Use
There are two NO2 filters in the LungFit PH; the main 12 hr NO2 filter (p/n 90008) that the user can change and the internal bagging 48 hr NO2 filter p/n 90009). The purpose of the NO2 filters is to remove NO2 from the NO gas flow and to collect any particulates that may be in the gas flow path. When the main 12 hr filter is inserted, this NO2 filter communicates with the LungFit device through an RFID wireless chip that is mounted on the filter. The RFID chip on the filter tracks and updates with the filter’s remaining usage time to prevent the filter from being accidently re-used after it has been consumed. The RFID also communicates wirelessly with the Gas Monitoring CPU, as described below. The NO2 filter includes two 1µm filters and two 5µm filters, with the NO2 filter material Sofnolime (soda lime) sandwiched between the particulate filters. Soda lime is a material that is commonly used in anesthesia CO2 absorbers. The duration of use for each main filter is pre-established as 12 hours. The RFID tag tracks filter usage so that the remaining time is updated continually every minute and displayed on the device user interface (i.e., as Time Remaining). The Gas Monitoring CPU detects whether a filter is installed and displays the Change Filter alarm to alert the user to change/insert a new user when no NO2 filter is present. The device also displays the Change Filter alarm when the NO2 filter has been consumed (no more minutes remaining).
CLOSE

Device Characteristics

MR Unsafe
No
No
Yes
Yes
No
No
No
No

GMDN

[?]

GMDN© Term Code, Names and Definitions (*Terms of Use): GMDN® is a registered trademark of The GMDN Agency. All rights reserved. Used under licence from The GMDN Agency Ltd.

GMDN Term Code GMDN Term Name GMDN Term Definition GMDN Term Status [?] Implantable?
37266 Nitric oxide delivery unit, system-based
A mains electricity (AC-powered) device, which may include internal rechargeable batteries, intended for the delivery of precise amounts of nitric oxide (NO), also known as nitrogen monoxide, to the respiratory tract of neonate, paediatric, and adult patients to treat severe respiratory disorders [e.g., primary pulmonary hypertension (PPH), acute respiratory distress syndrome (ARDS)]. It consists of a portable main unit that enables the delivery and monitoring of NO to gases that are to be breathed by the patient via a ventilator or other respiratory device/system. It typically includes accessory items (e.g., tubing, filters) and possibly a trolley (cart) for mobility.
Active false
CLOSE

FDA Product Code

[?]
Product Code Product Code Name
QTB Nitric Oxide Generator And Delivery System
CLOSE

FDA Premarket Submission

FDA Premarket Submission Number [?] Supplement Number [?]
P200044 000
No CLOSE

Sterilization

No
No
Sterilization Method [?]
No Sterilization Methods Found
CLOSE

Storage and Handling

[?]
Storage and Handling
No storage/handling found
CLOSE

Clinically Relevant Size

[?]
Size Type Text
No Device Sizes
CLOSE

Device Record Status

8011dee7-6f57-49fb-9467-7bb22a34437c
February 14, 2025
2
August 04, 2022
CLOSE

Alternative and Additional Identifiers Additional Identifiers

Package DI

[?]
Package DI Number Quantity per Package Contains DI Package Package Discontinue Date Package Status Package Type
B882900081 10 B882900080 In Commercial Distribution Box
B882900082 12 B882900081 In Commercial Distribution Box
CLOSE

Secondary DI

[?]
Issuing Agency [?] Secondary DI Number
No Secondary DIs found
CLOSE

Unit of Use DI

[?]
Unit of Use DI Number: No Unit of Use DI Numbers Found CLOSE

Direct Marking (DM)

[?]
No
No
None
CLOSE

Production Identifier(s) in UDI

[?]
Yes
No
Yes
Yes
No CLOSE

Customer Contact

[?]
1-855-586-4359
xxx@xx.xxx
CLOSE