The Corvis® ST is a non-contact tonometer equipped with an optical
Pachymetry function.
The Corvis® ST measures intraocular pressure without contact with the eye
by applying an air puff to the eye. During the air puff the eye gets
illuminated by a 9 mm slit through the apex and a built-in high-speed
camera records the movement of the eye with more than 4000 images per
second.
The high-speed camera uses a sequence of 140 Scheimpflug images of the
cornea which are analyzed by a built-in computer.
Intra ocular pressure is determined by the detection of the applanation
moments of the cornea.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
72300
In Commercial Distribution
- 04049584014085 ()
- Weight: 14 Kilogram
- Height: 520 Millimeter
- Depth: 520 Millimeter
- Width: 270 Millimeter
- Ophthalmic tonometer, battery-operated
The Pentacam Scheimpflug Camera is a non-invasive, diagnostic system created to take
photographs of the anterior segment of the eye, table mounted and AC powered. The system
is based on the Scheimpflug Principle for Slit Image photography. The device consists of a
measurement unit, power supply and a CPU. The measuring system uses blue light (UV-free)
given to a slit to illuminate the eye, and a CCD-Camera for photography. The measuring
system offers the possibility of automatically rotation to get photographs of every part of the
eye. The system calculates from the photos a 3D-model1 of the eye
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
70900
In Commercial Distribution
- 04049584000026 ()
- Height: 535 Millimeter
- Depth: 360 Millimeter
- Width: 280 Millimeter
- Weight: 11 Kilogram
- Corneal topography system
The Pentacam Scheimpflug Camera is a non-invasive, diagnostic system created to take
photographs of the anterior segment of the eye, table mounted and AC powered. The system
is based on the Scheimpflug Principle for Slit Image photography. The device consists of a
measurement unit, power supply and a CPU. The measuring system uses blue light (UV-free)
given to a slit to illuminate the eye, and a CCD-Camera for photography. The measuring
system offers the possibility of automatically rotation to get photographs of every part of the
eye. The system calculates from the photos a 3D-model1 of the eye.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
70700
In Commercial Distribution
- 04049584000019 ()
- Weight: 10.5 Kilogram
- Height: 535 Millimeter
- Depth: 360 Millimeter
- Width: 280 Millimeter
- Corneal topography system
While rotating around the eye, the Pentacam® AXL captures Scheimpflug
images of the anterior eye segment through varying axes. The
Scheimpflug images created during an examination are transmitted to
the connected PC.
The axial length of the eye is measured and displayed by interferometry.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
70100
In Commercial Distribution
- 04049584012333 ()
- Weight: 7.9 Kilogram
- Height: 430 Millimeter
- Depth: 208 Millimeter
- Width: 270 Millimeter
- Corneal topography system
The OCULUS Keratograph 5M combines the keratometric measuring process
with topographic mapping.
Measurement of the corneal surface is done by means of a Placido ring
system that is reflected off the cornea. These data are analyzed by the
computer.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
7700
In Commercial Distribution
- 04049584000477 ()
- Weight: 6.1 Kilogram
- Height: 512 Millimeter
- Depth: 400 Millimeter
- Width: 275 Millimeter
- Corneal topography system
The PARK 1 combines three measuring functions in one unit.
Auto-Refractometer
An infrared light source projects measuring light onto the retina of the eye
from where it is reflected back to the shutter location. Sensitive sensor
chips, or CCD cameras now register the deviation of the reflected light from
the shutter location. The deviation depends on the ametropia. From that, an
integrated microcomputer calculates the ametropia in D, based on the
sphere, cylinder and cylinder axis position.
1 Measuring ocular / Patient eyepiece 4 Chin rest
2 Keratometer ring 5 Marking for the eye height
3 Pachycam camera glass cover
Keratometer
To determine the curvature of the cornea, a reflected image of the cornea
is captured by a camera sensor and is measured.
The reflection of test marks and of a ring is used as the reflected image.
This allows the central radii of the cornea to be determined.
Pachymeter
The pachymetry principle uses Scheimpflug images of the cornea, which
are analysed by a built-in computer.
600 Absolute data points are evaluated with the Scheimpflug image. The
measuring range lies on a 4 mm slit through the apex.
The slit light illuminates a sectional plane from the front surface of the cornea
to the back surface. The transparent cells of the cornea scatter the slit
light such that the sectional plane appears as if it were self-luminous.
This is captured at an angle of 45° through the pupil by a camera, whereby
the image plane of the camera is also tilted 45° to the optical axis of the
camera lens, in order to sharply focus the light-scattering cornea plane
onto the image plane of the camera (Scheimpflug image).
Thanks to this arrangement, sharp sectional images of the cornea can be attained.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
65000
In Commercial Distribution
- 04049584000217 ()
- Weight: 12 Kilogram
- Depth: 530 Millimeter
- Width: 538 Millimeter
- Height: 265 Millimeter
- Ophthalmic refractometer, manual
The OCULUS Easyfield® has been designed for combined use as a
screening unit with all associated options for immediate follow-up
examination of suspicious findings. The most common examination grids
and strategies for the central visual field up to 30° can be accessed.
The Easyfield® offers pre programmed combinations for often needed
examination routines. For example: Screening 24-2, SPARK Quick,
Macula. You can also combine your own routines and then save them as
a program.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
15005
In Commercial Distribution
- 04049584000064 ()
- Weight: 4.6 Kilogram
- Height: 435 Millimeter
- Depth: 450 Millimeter
- Width: 275 Millimeter
- Perimeter, automatic
The OCULUS Easyfield® has been designed for combined use as a
screening unit with all associated options for immediate follow-up
examination of suspicious findings. The most common examination grids
and strategies for the central visual field up to 30° can be accessed.
The Easyfield® offers pre programmed combinations for often needed
examination routines. For example: Screening 24-2, SPARK Quick,
Macula. You can also combine your own routines and then save them as
a program.
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
15000
In Commercial Distribution
- 04049584000057 ()
- Weight: 6.8 Kilogram
- Height: 435 Millimeter
- Depth: 540 Millimeter
- Width: 316 Millimeter
- Perimeter, automatic
GONIO ready® 175 for f=175
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
10011228
In Commercial Distribution
- 04049584034489 ()
- 04049584034472 ()
- Microscope objective
GONIO ready® 200 for f=200
OCULUS Optikgeräte Gesellschaft mit beschränkter Haftung
10008422
In Commercial Distribution
- 04049584033048 ()
- 04049584034465 ()
- Microscope objective